Watson-Marlow Bredel e-manuals

  m-323dz-gb-03


Watson-Marlow 323Dz pumps


Clickable index

Contents

  1. Declaration of conformity
  2. Declaration of incorporation
  3. Warranty
  4. When you unpack your pump
  5. Information for returning pumps
  6. Peristaltic pumps - an overview
  7. Safety notes
  8. Specifications
  9. Good pump installation practice
  10. Care and maintenance
  11. Connecting to a power supply
  12. Start-up check list
  13. Switching on for the first time
  14. Switching on in subsequent power cycles
  15. Auto-restart
  16. The main menu

  1. Troubleshooting
  2. Part numbers
  3. 323S spares
  4. 323 Accessories
  5. 313D and 314D pumpheads
  6. 313/314 Flow rates
  7. Tubing for 313D and 314D pumpheads, 1.6mm wall
  8. Tubing for 313D2 and 314D2 pumpheads, 2.4mm wall
  9. 501RL pumphead
  10. 501 Flow rates
  11. Tubing for 501RL pumpheads, 1.6mm wall
  12. Trademarks
  13. Patient-connected applications
  14. Publication history
  15. Decontamination certificate

Double clicking anywhere in the manual will take you back to this index.

1 Declaration of conformity

This declaration was issued for Watson-Marlow 323Dz pumps on December 1, 2003. When this pump unit is used as a stand-alone pump it complies with: Machinery Directive 98/37/EEC BS EN 60204-1, Low Voltage Directive 73/23/EEC BS EN 61010-1, EMC Directive 89/336/EEC, BS EN 50081-1/BS EN 50082-1.

2 Declaration of incorporation

When this pump unit is to be installed into a machine or is to be assembled with other machines for installations, it must not be put into service until the relevant machinery has been declared in conformity with the Machinery Directive 98/37/EEC BS EN 60204-1.

Responsible person: Christopher Gadsden, Managing Director, Watson-Marlow Limited, Falmouth, Cornwall TR11 4RU, England. Telephone +44 (0) 1326 370370 Fax +44 (0) 1326 376009.

The information in this user guide is believed to be correct at the time of publication. However, Watson-Marlow Limited accepts no liability for errors or omissions. Watson-Marlow Bredel has a policy of continuous product improvement, and reserves the right to alter specifications without notice. This manual is intended for use only with the pump it was issued with. Earlier or later models may differ. The most up-to-date manuals appear on the Watson-Marlow website: http://www.watson-marlow.com

3 Two year warranty

Watson-Marlow Limited warrants, subject to the conditions below, through either Watson-Marlow Limited, its subsidiaries, or its authorised distributors, to repair or replace free of charge, any part of this product which fails within two years of delivery of the product to the end user. Such failure must have occurred because of defect in material or workmanship and not as a result of operation of the product other than in accordance with the instructions given in this manual.

Conditions of and specific exceptions to the above warranty are:

  • Consumable items such as tubing and rollers are excluded.
  • Products must be returned by pre-arrangement carriage paid to Watson-Marlow Limited, its subsidiaries, or its authorised distributor.
  • All repairs or modifications must have been made by Watson-Marlow Limited, its subsidiaries or its authorised distributors or with the express permission of Watson-Marlow Limited, its subsidiaries, or its authorised distributors.
  • Products which have been abused, misused, or subjected to malicious or accidental damage or electrical surge are excluded.

Warranties purporting to be on behalf of Watson-Marlow Limited made by any person, including representatives of Watson-Marlow Limited, its subsidiaries, or its distributors, which do not accord with the terms of this warranty shall not be binding upon Watson-Marlow Limited unless expressly approved in writing by a director or manager of Watson-Marlow Limited.

4 When you unpack your pump

Unpack all parts carefully, retaining the packaging until you are sure all components are present and in good order. Check against the components supplied lists, below.

Packaging disposal

Dispose of packaging materials safely, and in accordance with regulations in your area. Pay particular attention to the expanded polystyrene shockproof shells. The outer carton is made of corrugated cardboard and can be recycled.

Inspection

Check that all components are present. Inspect components for damage in transit. If anything is missing or damaged, contact your distributor immediately.

Components supplied (323Dz pump, standard product)

  • The 323Dz drive unit fitted with:
    • 501RL, 313 or 314 pumphead (see Pump specifications) if specified as a complete pump assembly
    • The designated mains power lead for your pump
  • PC-readable CDROM containing these operating instructions
  • Quick Start manual

Note: Some versions of this product will include components different from those listed above. Check against your purchase order.

Storage

This product has an extended shelf life. However, care should be taken after storage to ensure that all parts function correctly. Users should be aware that the pump contains a battery with an unused life of seven years. Long-term storage is not recommended for peristaltic pump tubing. Please observe the storage recommendations and use-by dates which apply to tubing you may wish to bring into service after storage.

5 Information for returning pumps

Equipment which has been contaminated with, or exposed to, body fluids, toxic chemicals or any other substance hazardous to health must be decontaminated before it is returned to Watson-Marlow or its distributor.

A certificate included at the rear of these operating instructions, or signed statement, must be attached to the outside of the shipping carton. This certificate is required even if the pump is unused.

If the pump has been used, the fluids that have been in contact with the pump and the cleaning procedure must be specified along with a statement that the equipment has been decontaminated.

6 Peristaltic pumps - an overview

Peristaltic pumps are the simplest possible pump, with no valves, seals or glands to clog or corrode. The fluid contacts only the bore of a tube, eliminating the risk of the pump contaminating the fluid, or the fluid contaminating the pump. Peristaltic pumps can operate dry without risk.

How they work

A compressible tube is squeezed between a roller and a track along an arc of a circle, creating a seal at the point of contact. As the roller advances along the tube, the seal also advances. After the roller has passed, the tube returns to its original shape, creating a partial vacuum which is filled by fluid drawn from the inlet port.

Before the roller reaches the end of the track, a second roller compresses the tube at the start of the track, isolating a packet of fluid between the compression points. As the first roller leaves the track, the second continues to advance, expelling the packet of fluid through the pump's discharge port. At the same time, a new partial vacuum is created behind the second roller into which more fluid is drawn from the inlet port.

Backflow and siphoning do not occur, and the pump effectively seals the tube when it is inactive. No valves are needed.

The principle may be demonstrated by squeezing a soft tube between thumb and finger and sliding it along: fluid is expelled from one end of the tube while more is drawn in at the other.

Animal digestive tracts function in a similar way.

Suitable applications

Peristaltic pumping is ideal for most fluids, including viscous, shear-sensitive, corrosive and abrasive fluids, and those containing suspended solids. They are especially useful for pumping operations where hygiene is important.

Peristaltic pumps operate on the positive displacement principle. They are particularly suitable for metering, dosing and dispensing applications. Pumps are easy to install, simple to operate and inexpensive to maintain.

7 Safety notes

In the interests of safety, this pump and the tubing selected should only be used by competent, suitably trained personnel after they have read and understood this manual, and considered any hazard involved. If the pump is used in a manner not specified by Watson-Marlow Ltd, the protection provided by the pump may be impaired.

Any person who is involved in the installation or maintenance of this equipment should be fully competent to carry out the work. In the UK this person should also be familiar with the Health and Safety at Work Act 1974.

This symbol, used on the pump and in this manual, means: Caution, risk of electric shock.

This symbol, used on the pump and in this manual, means: Caution, refer to accompanying documents.

This symbol, used in on the pump and this manual, means: Do not allow fingers to contact moving parts.

Fundamental work with regard to lifting, transportation, installation, starting-up, maintenance and repair should be performed by qualified personnel only. The unit must be isolated from mains power while work is being carried out. The motor must be secured against accidental start-up.

There are no user serviceable parts inside this pump. The unit should be returned to Watson-Marlow for service.

There are moving parts inside the pumphead. Before opening, ensure that the following safety directions are followed.

  • Ensure that the pump is isolated from the mains power.
  • Ensure that there is no pressure in the pipeline.
  • If a tube failure has occurred, ensure that any fluid in the pumphead has been allowed to drain to a suitable vessel, container or drain.
  • Ensure that protective clothing and eye protection are worn if hazardous fluids are pumped.
  • Primary operator protection from rotating parts of the pump is provided by the pumphead safeguard. Note that safeguards differ, depending on the type of pumphead. See the 313D and 314D pumpheads and the 501RL pumphead sections of this manual.

This pump must be used only for its intended purpose.

The pump must be accessible at all times to facilitate operation and maintenance. Access points must not be obstructed or blocked. Do not fit any devices to the drive unit other than those tested and approved by Watson-Marlow. Doing so could lead to injury to persons or damage to property for which no liability can be accepted.

If hazardous fluids are to be pumped, safety procedures specific to the particular fluid and application must be put in place to protect against injury to persons.

This product does not comply with the ATEX directive and must not be used in explosive atmospheres.

8 Pump specifications

Labels fixed to the rear of the pump contain manufacturer and contact details, product reference number, serial number and model details.

323Dz dispensing pump

This pump can be controlled from the keypad (or a footswitch for dosing). It features:

Manual control
Speed adjustment; run and stop; direction control.
Remote control
Dose starts can be controlled with a contact closure or logic input signal.
Dispensing
The pump dispenses measured volumes of fluid as single doses, batches or as a timed sequence of doses. Ramp and drip control.

High-speed model

323Dz with 313D three-roller flip-top pumphead: maximum speed 400 rpm; or 314D four-roller flip-top pumphead: maximum speed 300 rpm; for 1.6mm wall thickness tubing up to 8mm diameter.

Low-speed model

323Dz with 501RL 1.6mm wall tubing pumphead for tubing up to 8mm diameter: max speed 300 rpm. Note: It is possible to fit 313 and 314 pumpheads to a low-speed 323Dz drive by the use of an adaptor plate: part no 039.0031.000.

Power supply

115VAC/1ph/50/60Hz


230VAC/1ph/50/60Hz

Maximum rotor speed

300 rpm; 400 rpm

Speed control range

300:1; 200:1

Minimum rotor speed

1 rpm; 2 rpm

Power consumption

100VA

Operating temperature range

4C to 40C, 39F to 104F

Storage temperature range

-40C to 70C, -40F to 158F

Weight of drive

4.5kg, 9lb 15oz

Weight with 313 pumphead

4.8kg, 10lb 9oz

Weight with 501RL pumphead

5.5kg, 12lb 2oz

Noise

<70dB(A) at 1m

Standards

EN 60529-1 (IP31)


Machinery directive:
98/37/EEC BS EN 60204-1


Low voltage directive:
73/23/EEC BS EN 61010-1


EMC directive:
89/336/EEC BS EN 50081-1/BS EN 50082-1

Absolute minimum dose size

2 revolutions

Recommended minimum dose size

5 revolutions

Maximum single dose size

9.999litres

Minimum number of doses

1

Maximum number of doses

9,999

Minimum interval between doses

0.1 second

Maximum interval between doses

999 seconds

8.1 Dimensions

323Dz model

9 Good pump installation practice

9.1 General recommendations

A correctly engineered installation will promote long tube life. Site the pump on a flat, horizontal, rigid surface, free from excessive vibration, to ensure correct lubrication of the gearbox. Allow a free flow of air around the pump to ensure that heat can be dissipated. Ensure that the ambient temperature around the pump does not exceed 40C.

The STOP key on the keypad will always stop the pump. However, it is recommended that a suitable local emergency stop device is fitted into the mains supply to the pump.

Do not stack pumps more than three high. When pumps are stacked, ensure that the ambient temperature around all the pumps in the stack does not exceed 40C.

The pump may be set up so that the direction of rotor rotation is clockwise or counter-clockwise, whichever is convenient. Please note, however, that with a 501RL pumphead tube life will be greater if the rotor rotates clockwise; and that performance against pressure will be maximised if the rotor rotates counter-clockwise.

Peristaltic pumps are self-priming and self-sealing against backflow. No valves are required in inlet or discharge lines. Valves in the process flow must be opened before the pump operates. Users are advised to fit a pressure relief device between the pump and any valve on the discharge side of the pump to protect against damage caused by accidental operation with the discharge valve closed.

9.2 Do's and do not's

Do not build a pump into a tight location without adequate airflow around the pump.

Do keep delivery and suction tubes as short and direct as possible - though ideally not shorter than 1m - and follow the straightest route. Use bends of large radius: at least four times the tubing diameter. Ensure that connecting pipework and fittings are suitably rated to handle the predicted pipeline pressure. Avoid pipe reducers and lengths of smaller bore tubing than the pumphead section, particularly in pipelines on the suction side. Any valves in the pipeline (not usually needed with a self-priming peristaltic pump) must not restrict the flow. Any valves in the flow line must be open when the pump is running.

Do use suction and delivery pipes equal to or larger than the bore of the tube in the pumphead. When pumping viscous fluids use pipe runs with a bore several times larger than the pump tube.

Do ensure that on longer tube runs at least 1m of smooth bore flexible tubing is connected to the inlet and discharge port of the pumphead to help to minimise impulse losses and pulsation in the pipeline. This is especially important with viscous fluids and when connecting to rigid pipework.

Do site the pump at or just below the level of the fluid to be pumped if possible. This will ensure flooded suction and maximum pumping efficiency.

Do keep the pumphead track and all moving parts clean and free from contamination and debris.

Do run at slow speed when pumping viscous fluids. Flooded suction will enhance pumping performance in all cases, particularly for materials of a viscous nature.

Do re-calibrate after changing pump tubes, fluid, or any connecting pipework. It is also recommended that the pump is re-calibrated periodically to maintain accuracy.

IP31 models may be wiped down with a wet cloth, but should not be hosed or immersed.

When using Marprene or Bioprene continuous tubing, do re-tension the tube after the first 30 minutes of running.

Tube selection: The chemical compatibility lists published in Watson-Marlow publications are guides. If in doubt about the compatibility of a tube material and the duty fluid, request a Watson-Marlow tube sample card for immersion trials.

10 Care and maintenance

The pump is sealed to IP31 and is suitable for wipe-down cleaning. Do not use solvents, mechanical scourers, strong organic acids, or alkali-based cleaning solutions.

Remove any tubing, detach the pumphead, and wash the pumphead thoroughly with a mild solution of detergent in water.

Check moving parts of the rotor from time to time for freedom of movement. Lubricate pivot points and rollers occasionally with Teflon lubricating oil.

The pump has a good, broad chemical resistance to inorganic acids, saline solutions, alkalis, some hydrocarbons and a large number of oils and greases. It is suitable for wipe-down but not for long-term contact with alcohols. The case may be damaged by contact with strong acids or strong solvents

There are no user-serviceable parts inside the pump. The unit should be returned to Watson-Marlow or its appointed agents or distributors for service.

11 Connecting this product to a power supply

We recommend using commercially available supply voltage surge suppression where there is excessive electrical noise.

The voltage selector is located at the rear of the pump. Set the voltage selector to 115V for 100-120V 50/60Hz supplies or 230V for 200-240V 50/60Hz supplies. Always check the voltage selector switch before connecting the mains supply.

Make suitable connection to a single-phase mains electricity supply, using the supplied power cable with moulded plug. The cable uses the colour code:

  • 220-240V: live: brown; neutral: blue; earth: green/yellow.
  • 110-120V: live: black; neutral: white; earth: green.

Input line fusing: type T1,0A time-delayed fuse, located in the combined inlet socket and fuseholder at the rear of the pump.

Power interruption: This pump has an auto-restart feature which, when active, will restore the pump to the operating state it was in when power was lost. See Auto-restart.

Stop / start power cycles: Do not power up/power down for more than 10 starts per hour, whether manually or by means of the auto-restart facility.

12 Start-up check list

Note: See also Tube loading, 313 and 314 and Tube loading, 501RL.

  • Ensure that proper connections are achieved between the pump tube and suction and discharge piping.
  • Ensure proper connection has been made to a suitable power supply.
  • Ensure that the recommendations in the section on Good pump installation practice are followed.

13 Switching the pump on for the first time

Note: This manual uses bold type to highlight the active option in menu screens: "English" in the first screen represented here. The active option appears on the pump display in inverse text.

  • Switch on the power supply at the rear of the pump. The pump runs a power-on test to confirm proper functioning of the memory and hardware. If a fault is found, an error message is displayed. See Error codes.
  • The pump displays a language menu. Use the UP and DOWN keys to select your language. Press the RETURN key to confirm your choice.
    The information which follows assumes that your choice was English.
    When the language is chosen this menu will not appear again and all menus will appear in the language you chose. (Language can be reset as described later. See Language.)
  • The pump displays the Watson-Marlow start-up screen for four seconds, followed by the pump model identity screen for four seconds (examples are shown here), and then the main menu.
  • The rotation symbol on the display indicates clockwise rotation. The speed of rotation is the pump's maximum. Initial start-up operational parameters are listed in the table below.

First-time start-up defaults

Language

Not set

Speed

Maximum

Direction

Clockwise

Calibration

400 rpm:
313, 8mm tube
300 rpm:
501RL, 8mm tube

Backlight

On

Auto-restart

Off

Pump status

Stopped

Beeper

On

The pump is now ready to operate according to the defaults listed above.

All operating parameters may be changed by means of key-presses. See Manual operation.

14 Switching the pump on in subsequent power cycles (if not in auto-restart mode)

  • Switch on the power supply at the rear of the pump. The pump runs a power-on test to confirm proper functioning of the memory and hardware. If a fault is found, an error message is displayed. See Error codes.
  • The pump displays the Watson-Marlow start-up screen for four seconds followed by the pump model identity screen for four seconds, and then the main menu.
    • Note: If ANY key is pressed during the display of any of the preliminary screens, the display jumps to the next screen. Quickly pressing any two keys or any key twice immediately after switch-on causes the display to jump to the main menu. In the main menu, keys assume their normal functions - see Manual operation below.
  • Start-up defaults are those in place when the pump was switched off last. Check that the pump is set to operate as you require it.

The pump is now ready to operate. If the pump starts immediately, look for the ! symbol on the display. This symbol indicates that the pump is set for auto-restart. Press the STOP key to stop the pump. See Auto-restart.

All operating parameters may be changed by means of key-presses. See Manual operation.

15 Auto-restart

Auto-restart will re-start the pump after mains power interruptions.

In auto-restart mode, if the pump was previously dispensing a dose, it returns to the dose start screen and waits for the START key (or remote dose start switch) to be pressed.

If the pump was previously under manual control, auto-restart returns the pump to the last manual setting; stopped if the pump was previously stopped, or running if the pump was previously running.

The default setting is auto-restart off. Without auto-restart the pump displays the main menu and waits for a control mode to be selected.

  • Mains power must be available to the pump to engage auto-restart.
  • Stop the pump. Turn off the mains power switch at the rear of the pump.
  • Hold down the START key and turn the mains power switch on. The ! symbol shows on the display.
  • Start the pump. If the mains supply is interrupted the pump will automatically re-start when the mains power returns.
  • Auto-restart is retained while the pump is switched off.
  • To remove auto-restart switch off the mains power at the rear of the pump. Hold down the STOP key and turn the mains power switch on. The ! symbol will go out.

Do not use auto-restart for more than 10 starts per hour.

16 The main menu

In addition to their functions in other operations (see Manual control), the following keys have specific actions in menu screens:

  • STOP: In general, STOP functions as a "go back" key, taking the user up one menu level without making a change. In a numeric entry screen, STOP resets the pump to the previous value.
  • UP: The UP key is used in menu item selection: it moves a highlight up the menu. When a numerical entry screen is displayed, pressing UP increases the number displayed.
  • DOWN: The DOWN key is used in menu item selection: it moves a highlight down a menu. When a numerical entry screen is displayed, pressing DOWN decreases the number displayed.
  • START: During Dose setup, press the START key to leave setup and begin the dose batch via the priming screen. When the priming screen is displayed, press and hold the START key to operate the pump at full speed and prime the pump.
  • DIRECTION: During Dose setup, the DIRECTION key reverses the direction of pump rotation instantly.
  • RETURN: The RETURN key functions in a similar way to the "enter" key of a personal computer: it confirms key-presses made immediately before. In menu item selection, it triggers the action or display selected from a menu using the UP and DOWN keys.

The options are Dose, Calibrate, Manual, Setup.

  • Dose programs the pump for dispensing.
  • Calibrate sets the pump for accurate flow. The flow rate of the pump is governed by the tubing. Factory default is for the 501 pumphead with 8mm bore tubing or the 313/314 pumphead with 8mm bore tubing.
  • Manual for keypad control of continuous fluid transfer.
  • Setup prepares the pump for operation.

Use the UP and DOWN keys to move between the menu options. Press the RETURN key to select the option.

Press the STOP key to exit a sub menu and move back up to the previous menu. Or press the STOP key, when changing values on the display, to revert to the original value.

16.1 Setup

Use the UP and DOWN keys to select Setup from the main menu. Press RETURN. The pump displays the setup menu: ROM, Backlight, Defaults, Language and Menu.

Scroll through the menu using the UP and DOWN keys. Press RETURN to select the option you wish to change.

ROM version

Use the UP and DOWN keys to select ROM. Press RETURN. The display shows the software version, the pump type, and drive speed for 4 seconds.

Display backlight

Use the UP and DOWN keys to select Backlight from the Setup menu. Press RETURN. Use the UP and DOWN keys to select backlight On or Off. Press RETURN.
Alternatively ...
When not in the Setup menu, hold down the STOP and UP keys to switch the backlight on; hold down the STOP and DOWN keys to switch the backlight off.

Restore factory defaults

This clears any programmed settings and restores the pump to the original factory settings. Use the UP and DOWN keys to select Defaults. Press RETURN. The pump displays a brief warning message that all settings will be deleted, and offers a choice: Restore defaults: Yes or No. Use the UP and DOWN keys to select Yes. Press RETURN. The pump displays the Setup menu.

Language

Use the UP and DOWN keys to select Language from the setup menu. Press RETURN. Use the UP and DOWN keys to select English, Français, Deutsch or Español. Press RETURN. The pump displays the setup menu in your chosen language. It is assumed for the rest of this manual that your choice was English.

Menu

To exit the Setup menu, and return to the main menu, use the UP and DOWN keys to select Menu. Press RETURN.
Alternatively ...
Press STOP.

16.2 Manual control

All settings and functions of the pump in manual mode are set and controlled by means of key-presses. The currently selected rotation direction is indicated on the display by a clockwise or counter-clockwise segmented arrow. If an exclamation mark ( ! ) shows, it indicates that Auto-restart is on.

Note: A number of the controls listed below are shortcuts to commands which are also available through the main menu. See Main menu.

A brief single press on each key triggers a beep sound (if enabled - see Keypress combinations, below) and causes the pump to function as follows:

  • START: starts the pump at the speed and in the direction shown on the display. The rotation symbol will become animated to confirm that the pump is operating. We recommend that the speed is reduced to a minimum before starting the pump.
  • STOP: has no effect if the pump is not running. If the pump is running, pressing STOP stops the pump. The display continues to show the previous speed and direction. The pump returns to this speed and direction when the START key is pressed again.
  • UP: increases the speed shown on the display in minimum steps of 1 rpm (unless the speed displayed is already the maximum speed). If the pump is then started by pressing the START key, it operates at the new speed. If the pump is running when UP in pressed, the change takes effect immediately.
  • DOWN: decreases the speed shown on the display in minimum steps of 1 rpm. If the pump is then started by pressing the START key, it operates at the new speed. The minimum speed possible is 1 rpm (low speed model) or 2 rpm (high speed model). If the pump is running when DOWN is pressed, the change takes effect immediately.
    Note: You can reduce the pump speed to 0 rpm by a further press on the DOWN key. The pump is still in the running state and the rotation symbol will continue to move. Press the UP key to return the pump to the minimum speed.
  • DIRECTION: toggles the direction of rotation shown on the display. If the pump is then started by pressing the START key, it rotates in the new direction. If the pump is running when DIRECTION is pressed, the change takes effect immediately.
  • RETURN: stops the pump (if it is running) and displays the main menu.

Keypress combinations cause the pump to function as follows:

  • UP and DIRECTION on power-up: toggles the keypad beep on and off.
  • START on power-up: switches on the Auto-restart facility. See Auto-restart.
  • STOP on power-up: switches off the Auto-restart facility. See Auto-restart.
  • STOP and UP: turns the display backlight on.
  • STOP and DOWN: turns the display backlight off.
  • DIRECTION and DOWN: interrupts the display to show the pump's ROM version for four seconds.

Use the UP and DOWN keys to select Manual from the main menu. Press RETURN. The display shows the last set speed. (An example is shown here.)

  • Use the UP key to increase the set speed. Use the DOWN key to reduce the set speed. We recommend that the speed is reduced to a minimum before starting the pump.
  • Press the DIRECTION key to reverse the direction of rotation.
  • The direction is shown by the rotation symbol. The direction may be changed while the pump is stopped or running.
  • Start the pump with the START key.
  • The rotation symbol moves to confirm that the pump is operating. The symbol is static when the pump is stopped.
  • Stop the pump with the STOP key. The pump stops immediately.
  • The display continues to show the previous speed and direction. The pump returns to this speed when the START key is pressed again.
  • You can reduce the pump speed to 0 rpm with the DOWN key. The pump is still in the running state and the rotation symbol continues to move. Press the UP key to return the pump to the minimum speed.

Press the RETURN key to return to the main menu. If the pump is operating it stops, and the main menu is displayed.

16.3 Calibration procedure

To dispense the correct amount of fluid the pump must know which pumphead is fitted and the tube size in the pumphead. You may also measure the flow from the pump and enter this value for the most accurate calibration.

  • Use the UP and DOWN keys to select Calibrate from the main menu. Press RETURN.
  • The pump shows the stored values of pumphead, tube size, and present flow rate in ml per minute. Example figures are shown here.
  • If the pumphead and tube information is correct, select Accept and press RETURN. The pump briefly displays a summary of the current speed and flow settings, and the main menu.
  • If the pumphead and tube information is wrong use the UP and DOWN keys to select Change and press RETURN.
    • The display lists pumphead options. Use the UP and DOWN keys to select the correct one and press RETURN.
    • The display lists the tube sizes. Use the UP and DOWN keys to select the correct size and press RETURN. The display asks the user if he wishes to run a calibration dose.

  • The display asks the user if he wishes to run a calibration dose. To obtain the most precise calibration, select Yes. Select No if you wish to use the pre-programmed flow data and return to the main menu via a summary of the current speed and flow settings. Press RETURN.
  • If you choose to run a calibration dose, the pump displays its current direction and rotation speed. You may change pump rotation direction and speed using the DIRECTION and UP and DOWN keys.
  • Put a measuring container at the pump outlet. Press START. The pump runs for 4 minutes, displaying an information screen for 15 seconds and a further information screen for the rest of the 4 minutes. You may stop the calibration dose at any time with the STOP key - but allow the pump to run as long as possible to obtain the most accurate calibration. A minimum of 15 seconds is recommended.
  • Measure the quantity of fluid dispensed. The measurements must be in millilitres for calibration and dosing. The pump displays its calculated dose volume, based on previous calibration data. Use the UP and DOWN keys to adjust this reading to match the measured volume. Press RETURN. The pump displays the new head, tube and flow settings, and redisplays the main menu. Example figures are shown here.

It is also possible to re-calibrate the pump during a batch run. (See Dose.) Re-calibration allows fine tuning of basic calibration (see Re-calibration in the Dose section).

If a re-calibration differs by more than 25% from the original calibration, the re-calibration value will be ignored. Another full calibration is required to change a dose size by more than 25%.

16.4 Dose

The Dose facility programs the pump to dispense measured volumes of fluid. These can be individual doses, a batch of doses delivered singly, or a batch of doses delivered at timed intervals. A dose, or batch of doses, can be controlled by the pump, by pressing the START key, or by pressing an optional remote-control foot or hand switch or by an external logic signal.

You can start pumping immediately using the settings chosen for the previous batch; or you can alter one or more of the settings before starting to pump.

To start pumping immediately:

  • Use the UP and DOWN keys to select Dose from the main menu. Press RETURN. The pump displays the dose size, dosing interval and the number of doses and waits for the START key to be pressed.
  • When the START key is pressed, the pump offers the user the opportunity to prime the pump. To do so, press and hold the START key. The pump operates continuously at maximum speed until the START key is released.
  • When priming is complete, or if no priming is required, press RETURN. The pump displays the dose size and the number of doses and waits for the START key to be pressed to start the batch.

To alter the batch settings before starting pumping:

  • Use the UP and DOWN keys to select Dose from the main menu. Press RETURN.
  • The pump displays the first three of the eight parameters which may be set for the next dosing session: dose volume, dosing interval and number of doses.

    The Dose menu occupies three screens.

    To move from one screen to subsequent screens, repeatedly press DOWN. Each item is highlighted in turn until the last item on the screen is highlighted. A further press on the DOWN key displays the next screen of the menu, with the first item highlighted.

    Follow the reverse procedure using the UP key to move to an item on a previous screen of the menu.

    Make a selection using the UP or DOWN keys and press ENTER to confirm your choice. The values shown are those set for the last dosing session. As each parameter of the eight is highlighted, you may accept or change its value.

  • If it is correct, do nothing. UP and DOWN will highlight the next parameter.
  • If you wish to change it, press RETURN. Use the UP and DOWN keys to alter the value. When the value is correct, press RETURN.
  • At any time during the dose setup sequence described below the user may press START and the pump displays the priming screen and may be operated according to the parameters in force. If the next batch requires only the dose size to be changed from the last batch, for example, change it and press START, ignoring the other seven parameters.

In this programming area, pressing STOP while setting up dosing parameters returns the value being changed to its original setting, allowing the user to start again.

Dose volume

  • Use the UP and DOWN keys to highlight the volume line of the display (the top line, first screen).
  • If the dose volume displayed is correct, do nothing. UP and DOWN highlight the next or previous parameter.
  • If you wish to change the dose volume, press RETURN and use the UP and DOWN keys to scroll the display to the required dose volume. Example figures are shown here. To quickly return to the starting value, press STOP. When the dose volume is correct, press RETURN. The first screen of the dose menu is redisplayed with the new dose size. If you wish to change other parameters, use the UP and DOWN keys to highlight each as required.

Note: The dose size must use more than two complete revolutions of the pumphead. To obtain best accuracy the dose size should use more than five revolutions. Where the dose size uses less than five revolutions, the pump may warn that a smaller tube size is required. If you do not change the tube size, the pump may proceed with your selected dose size but accuracy may be reduced. If the dose is smaller than the two revolution minimum, the pump displays a warning and does not allow you to proceed. If you wish to proceed with the dose size, you must perform a new pump calibration using a smaller tube size. (See Calibration).

Dose interval

  • Use the UP and DOWN keys to highlight the dose interval line of the display (the second line, first screen). The display shows the previously set time interval between doses in seconds.
    Note: If the time is set to zero, the pump waits for a start signal from the START key, or from an optional external control switch or logic signal, before proceeding with each dose. If the time interval is greater than zero, the pump proceeds through the sequence of doses at the programmed time interval.
  • If the interval displayed is correct, do nothing. UP and DOWN highlight the next or previous parameter.
  • If you wish to change the interval between doses, press RETURN and use the UP and DOWN keys to scroll the display to the required interval. Example figures are shown here. To quickly return to the starting value, press STOP. When the dose interval is correct, press RETURN. The first screen of the dose menu is redisplayed with the new dose interval. If you wish to change other parameters, use the UP and DOWN keys to highlight each as required.

Number of doses

  • Use the UP and DOWN keys to highlight the dose number line of the display (the third line, first screen). The display shows the previously set number of doses.
  • If the number displayed is correct, do nothing. UP and DOWN highlight the next or previous parameter.
  • If you wish to change the number of doses, press RETURN and use the UP and DOWN keys to scroll the display to the required number. Example figures are shown here. To quickly return to the starting value, press STOP. When the number is correct, press RETURN. The first screen of the dose menu is redisplayed with the new number of doses. If you wish to change other parameters, use the UP and DOWN keys to highlight each as required.

Note: If the number of doses is set to 1, the pump waits for a press on START for each dose, and the display during dosing increments with each dose. If the number of doses is set to greater than 1, the display decrements with each dose until it reads zero and the batch is complete.

Pump speed

  • Use the UP and DOWN keys to highlight the speed line of the display (the first line, second screen). The display shows the previously set pump speed.
  • If the speed displayed is correct, do nothing. UP and DOWN will highlight the next or previous parameter.
  • If you wish to change the speed, press RETURN and use the UP and DOWN keys to scroll the display to the required speed, up to a maximum of 300 rpm (low-speed model or high-speed model fitted with a 314 pumphead) or 400 rpm (high-speed model fitted with a 313 pumphead). Example figures are shown here. To quickly return to the starting value, press STOP. When the speed is correct, press RETURN. The second screen of the dose menu is redisplayed with the new pump speed. If you wish to change other parameters, use the UP and DOWN keys to highlight each as required.

Pump direction

  • Use the UP and DOWN keys to highlight the direction line of the display (the second line, second screen). The display shows the previously set direction of rotation.
  • If the direction displayed is correct, do nothing. UP and DOWN will highlight the next or previous parameter.
  • If you wish to change the direction, press RETURN and use the UP or DOWN keys to highlight the desired direction. An example is shown here. Press RETURN. The second screen of the dose menu is redisplayed with the new pump direction. If you wish to change other parameters, use the UP and DOWN keys to highlight each as required.

Alternatively ...
Press the DIRECTION key at any point in the menu sequence.

Start ramp

  • Use the UP and DOWN keys to highlight the start ramp line of the display (the third line, second screen). The display shows the start ramp setting of the last dosing session. When set to ‘0’ the pump starts abruptly at full speed. When set to ‘5’ the pump accelerates softly to full speed.
  • If the setting displayed is correct, do nothing. UP and DOWN will highlight the next or previous parameter.
  • If you wish to change the setting, press RETURN and use the UP or DOWN keys to highlight the desired setting: 0, 1, 2, 3, 4 or 5. An example is shown here. To quickly return to the starting value, press STOP. Press RETURN when correct. The second screen of the dose menu is redisplayed with the new start ramp setting. If you wish to change other parameters, use the UP and DOWN keys to highlight each as required.

End ramp

  • Use the UP and DOWN keys to highlight the end ramp line of the display (the first line, third screen). The display shows the end ramp setting of the last dosing session. When set to ‘0’ the pump stops abruptly. When set to ‘5’ the pump decelerates softly to a stop.
  • If the setting displayed is correct, do nothing. UP and DOWN will highlight the next or previous parameter.
  • If you wish to change the setting, press RETURN and use the UP and DOWN keys to highlight the desired setting: 0, 1, 2, 3, 4 or 5. An example is shown here. To quickly return to the starting value, press STOP. Press RETURN when correct. The third screen of the dose menu is redisplayed with the new end ramp setting. If you wish to change other parameters, use the UP and DOWN keys to highlight each as required.

Drip

The pump can be set to reverse briefly at the end of each dose to prevent drips.

  • Use the UP and DOWN keys to highlight the drip line of the display (the second line, third screen). The display shows the drip setting of the last dosing session: from 0 to 1.0: zero to one revolution at one-tenth of a revolution intervals.
  • If the setting displayed is correct, do nothing. UP and DOWN will highlight the next or previous parameter.
  • If you wish to change the setting, press RETURN and use the UP and DOWN keys to increase or decrease drip reversal: zero revolutions to 1 revolution at 0.1 revolution intervals. An example is shown here. To quickly return to the starting value, press STOP. Press RETURN when correct. The third screen of the dose menu is redisplayed with the new drip setting. If you wish to change other parameters, use the UP and DOWN keys to highlight each as required.

Proceed

The third line of the third screen presents the option Proceed.

  • If you wish to begin the dosing session, use the UP or DOWN keys to highlight Proceed and press RETURN. The pump offers the user the opportunity to prime the pump. To do so, press and hold the START key. The pump operates continuously at maximum speed until the START key is released. When priming is complete, or if no priming is required, press RETURN. The pump displays the dose size and the number of doses and waits for the START key to be pressed.
  • If you wish to review or change the parameters for the next batch, press STOP twice and use the UP and DOWN keys to move back up through the list of parameters.

Single dose dispensing

When START is pressed, the pump performs a dosing batch according to the parameters set.

If the batch is set to perform one dose, as in the example shown here, the pump stops when that dose has been dispensed and waits for a further press on the START key. As each dose is dispensed, the display increments, recording the number of doses dispensed so far.

Note: To dispense a batch of one dose and set the pump to count the doses/batches, check that the dose interval is set to zero seconds. If an interval is set, the pump will still dispense batches of one dose, but will not increment its display to show how many doses/batches have been dispensed.

Batch dispensing

If the batch is set to perform more than one dose with a set interval between doses, the pump operates until all the batch's doses have been dispensed. The display decrements as each dose is dispensed, showing the number of doses still to be dispensed. During the intervals between doses - 3 seconds in the example shown - the time elapsing is counted down to show how much time is left before the next dose will start.

If the interval is set to zero seconds, the batch can be dispensed in individually triggered doses by repeated presses on the START key. The display decrements as each dose is dispensed, showing the number of doses still to be dispensed.

When the batch is complete, the pump displays a screen summarising the batch, and offering two choices: start the batch again or return to the menu screen, where batch setting changes can be made.

  • Use the UP and DOWN keys to highlight Start or Menu and press RETURN.

  • If you select Start, the pump displays the dose size and the number of doses and waits for the START key to be pressed to begin a repeat batch.
    Alternatively ...
    Just press START.
    Note: When a batch is repeated, no opportunity is offered to prime the pump. If you wish to re-prime the pump, return to the main menu and re-enter the dosing menu as if it were a new batch.

  • If you select Menu, the pump displays the main menu. See Main menu.

Batch pause

You can interrupt a batch at any time by pressing STOP. The pump stops and offers four choices: Unpause: continue the dose sequence from where it was stopped; Restart: start the dose or the batch from the beginning; re-calibrate: re-calibrate the pump; and Exit: abandon the batch and return to the main menu.

  • Use the UP and DOWN keys to make a selection. Press RETURN to confirm.
  • Unpause: If you choose Unpause the pump finishes dispensing the current dose and completes the batch.

  • Restart: If you choose Restart the pump displays the number of doses in the batch and the number which have been completed. It offers two choices: restart the interrupted dose or restart the batch. Use the UP and DOWN keys to make a selection. Press RETURN to confirm.

    • Select Restart dose and the pump re-dispenses the interrupted dose and resumes the batch where it was interrupted - as in the example shown above.

    • Select Restart batch to command the pump to display the size and number of doses in the batch, as in the example above, and wait for the START key to be pressed to restart the batch.

  • Re-calibrate: (Note: In this manual, re-calibrate means adjust the full calibration previously performed. It is not possible at this point to perform a fresh full calibration. To do so return to the main menu. See Calibration.) If you choose re-calibrate you can check a recent dose and adjust the dose size (within a range of ±25%) for the remainder of the batch.
  • The display shows the volume of fluid it believes it has dispensed in each dose.
  • Measure the volume of fluid dispensed in a recent dose.
    • If the pump's display is correct, press RETURN.
    • If the pump's display is incorrect, use the UP and DOWN keys to adjust the volume displayed to match the measured quantity of fluid. Press RETURN to leave the re-calibration sequence.
    • If the volume adjustment needed is greater than 25%, it is likely that the original calibration is incorrect. The pump displays a warning and resets to the original calibration value. You must perform a new pump calibration. Press RETURN to leave the re-calibration sequence.
    • The pump redisplays the four choices: Unpause; Restart; re-calibrate; and Exit.
    • Use the UP and DOWN keys to make a selection. Press RETURN to confirm.
    • If you choose Unpause or Restart the pump operates as described above, but using a dose adjusted as a result of any changes you made during the re-calibration sequence. If you choose re-calibrate, you can repeat the re-calibration. If the pump did not allow re-calibration because the adjustment required was greater than 25%, choose Exit. The pump displays the main menu, from which you can perform a fresh complete calibration of the pump.

  • Exit: If you choose Exit, the pump redisplays the main menu.

16.5 Remote control

A Watson-Marlow remote control footswitch or handswitch may be used to start the dose or batch. The dose will proceed once the switch is pressed. In an emergency press the STOP key to halt the dose.

The switch should be connected as shown. Or a TTL-compatible logic signal may be applied to pin 8. (Low 0V, High 5V maximum. Ground to pin 9).

Never apply mains voltage to the 25-way D socket. Up to 5V TTL may be applied across pins 8 and 9, but do not apply voltage across any other pins. It may result in permanent damage not covered by warranty.

17 Troubleshooting

If the pump display remains blank when the pump is switched on, make the following checks:

  • Check the position of the voltage selector switch. The voltage selector is mounted at the rear of the pump.
  • Check the mains power switch at the rear of the pump.
  • Check that mains power is available to the pump.
  • Check the fuse in the fuseholder at the rear of the pump.
  • Check the fuse in the wall plug if one is present.

If the pump runs but there is little or no flow, make the following checks:

  • Check that the tube and rotor are in the pumphead.
  • Check that fluid is supplied to the pump.
  • Check that the tube is not split or burst.
  • Check for any kinks or blockages in the lines.
  • Check that any valves in the lines are open.
  • Check that the correct wall-thickness tube is being used.
  • Check direction of rotation.

17.1 Error messages

Error

Error condition

Suggested action

0

RAM write error

Attempt to reset by switching power OFF / ON. Or seek support

1

RAM corruption

Attempt to reset by switching power OFF / ON. Or seek support

2

OTP ROM error / corruption

Attempt to reset by switching power OFF / ON. Or seek support

3

OTP ROM read error

Attempt to reset by switching power OFF / ON. Or seek support

5

Unknown pump type

Check the interface card and cables. Attempt to reset by switching power OFF / ON. Or seek support

7

Display failure

Call supplier

8

Wrong key-press

Attempt key-press again. Attempt to reset by switching power OFF / ON

9

Motor stalled

Stop pump immediately. Check pumphead and tube. Power OFF / ON may reset. Or seek support

10

Tacho fault

Stop pump immediately. Power OFF / ON may reset. Or seek support

14

Over speed

Stop pump immediately. Power OFF / ON may reset. Or seek support .

15

Over current

Stop pump immediately. Check system. Power OFF / ON may reset. Or seek support

16

Over voltage

Stop pump immediately. Check mains voltage selector switch. Check supply. Power OFF / ON may reset. Or seek support

17

Under voltage

Stop pump immediately. Check mains voltage selector switch. Check supply. ON / OFF may reset. Or seek support

18

Watchdog error

Attempt to reset by switching power OFF / ON. Or seek support

19

Over temperature

Stop pump immediately. Turn OFF. Call supplier

20

Signal out of range

Check analogue control signal range. Trim signal as required. Or seek support

21

Analogue over signal

Reduce the analogue control signal

22

No analogue signal

Connect analogue control signal or return to manual control

33

Unrecognised
key-press

Attempt key-press again. Attempt to reset by switching power OFF / ON. Or seek support

35

Work overload

Turn OFF. Check power supply. Check pumphead and tubing. Wait 30 minutes. Power ON may reset. Or seek support

Err

General error condition

Turn OFF. Call supplier

18 Part numbers

Drives only

Drive type

Drive speed

Pumphead

Mains lead

036.3183.00U

Dz

300

-

UK

036.3184.00U

Dz

400

-

UK

036.3183.00E

Dz

300

-

Euro

036.3184.00E

Dz

400

-

Euro

036.3183.00A

Dz

300

-

US

036.3184.00A

Dz

400

-

US

Complete pump assemblies

030.3183.RLU

Dz

300

501RL

UK

030.3184.3DU

Dz

400

313D

UK

030.3184.4DU

Dz

400

314D

UK

030.3183.RLE

Dz

300

501RL

Euro

030.3184.3DE

Dz

400

313D

Euro

030.3184.4DE

Dz

400

314D

Euro

030.3183.RLA

Dz

300

501RL

US

030.3184.3DA

Dz

400

313D

US

030.3184.4DA

Dz

400

314D

US

19 323Dz spares

1

FB0009

Foot

2

MN2094T

Interface card cover

3

FS0003

Fuse

20 323 Accessories

520AF

Footswitch

059.3002.000

520AH

Hand switch

059.3022.000

505AS

Filling stand

059.5001.000

505AL

Dispensing lance

059.5052.000

Tube life

Application factors that influence tube life in peristaltic pumps are pump speed and number of rollers (roller impacts/minute), chemical compatibility and viscosity of the duty fluid and suction and discharge pressure. Several tubing materials are available with different levels of life expectancy. In addition, tubing's dimensional tolerances from the manufacturing process also affect its life.

In perfect conditions of no suction or discharge pressure, pumping water in a clean environment at normal room temperature, the following nominal tube life may be experienced:


313/314

501RL

Marprene, Bioprene,
StaPure and ChemSure

less than 6,000 hours

less than 10,000 hours

Silicone

less than 250 hours

less than 250 hours

Others

less than 100 hours

less than 100 hours

Note: This is a guide only. It is not possible to make an accurate assessment of tube life except by trial of a real application.

21 313D and 314D pumpheads

The 313D pumphead has three rollers and is designed to provide higher flow rates. The 314D pumphead has four rollers to provide greater pumping precision with less pulsation in the flow. Both designs are available for 1.6mm and 2.4mm wall tubes.

New tubing can be loaded easily into the flip-top design. The top closes with a "clamp and stretch" action to locate the tube in the correct position and with the correct tension.

Standard and extension pumpheads are bayonet mounted. This ensures easy cleaning and fast set up.

21.1 313D and 314D installation and removal

Always isolate the pump from the mains power supply before opening the guard or performing any positioning, removal or maintenance activity.

Installation

  • Engage the pumphead drive slot with the end of the pump drive shaft. Continue to align the pumphead until the bayonet engages with the mounting plate. Turn the pumphead clockwise until it locks into an upright position.

Removal

  • Push the locking lever back and turn the pumphead anti-clockwise until it is free from the mounting plate.

21.2 313D and 314D tube loading




  • Switch off the mains supply. Lift the flip-top track until it is fully open.
  • Set the tube clamps to the correct tube size. The track must be fully open. Align the scale on both sides of the pumphead. If the tube is slippery, or if there is a high suction lift, the tube clamps may need a smaller setting to secure the tube.

  • Select enough tube length for the curve of the pump track. Slide the tube into the open pumphead. The tube must not be twisted or stretched against the rollers.
  • Ensure the tubing locates in the centre of the tube clamps. Carefully lower the track. Check that the tube is not crushed in the clamps or over-stretched.

When using Marprene tubing

Re-tension new tube after the first 30 minutes of running. Stop the pump and release the flip top. Allow the tube to settle naturally across the rollers. Re-clamp the tube. Re-start the pump. This will correct the normal stretching that occurs with new Marprene tube. The correct tension is essential for good tube life.

21.3 313D and 314D pumphead spares

1

033.3411.000

313D three-roller pumphead

2

033.3431.000

313X extension three-roller pumphead

1

033.4411.000

314D four-roller pumphead

2

033.4431.000

314X four-roller extension pumphead

1

033.3511.000

313D2 three-roller pumphead for 2.4mm wall tube

2

033.3531.000

313X2 three-roller extension pumphead for 2.4mm wall tube

1

033.4511.000

314D2 four-roller pumphead for 2.4mm wall tube

2

033.4531.000

314X2 four roller extension pumphead for 2.4mm wall tube

22 313/314 Flow rates

Flow rates were obtained using silicone tubing with the pumphead rotating clockwise, pumping water at 20C with zero suction and delivery pressures. For critical applications determine flow rates under operating conditions.

22.1 313D 1.6mm wall: ml/min

313D

Speed
range

0.5mm
1/50in
# 112

0.8mm
1/32in
# 13

1.6mm
1/16in
# 14

3.2mm
1/8in
# 16

4.8mm
3/16in
# 25

6.4mm
1/4in
# 17

8.0mm
5/16in
No 18

2-400rpm

0.06-12

0.14-28

0.54-110

2-400

4.4-880

7.2-1400

10-2000

22.2 314D 1.6mm wall: ml/min

314D

Speed
range

0.5mm
1/50in
# 112

0.8mm
1/32in
# 13

1.6mm
1/16in
# 14

3.2mm
1/8in
# 16

4.8mm
3/16in
# 25

6.4mm
1/4in
# 17

8.0mm
5/16in
# 18

2-400rpm

0.06-12

0.12-24

0.50-100

1.7-340

3.8-760

6.0-1200

8.0-1600

22.3 Maximum number of pumpheads

313D, 314D, Pumpsil, to 400 rpm

0.5mm
1/50in
# 112

0.8mm
1/32in
# 13

1.6mm
1/16in
# 14

3.2mm
1/8in
# 16

4.8mm
3/16in
# 25

6.4mm
1/4in
# 17

8.0mm
5/16in
# 18

0-0.5 bar

6

6

5

3

2

2

1

0.5-2 bar

6

6

5

3

2

1

1

313D, 314D, Marprene, Bioprene, Tygon, Neoprene, Fluorel, to 400 rpm

0.5mm
1/50in
# 112

0.8mm
1/32in
# 13

1.6mm
1/16in
# 14

3.2mm
1/8in
# 16

4.8mm
3/16in
# 25

6.4mm
1/4in
# 17

8.0mm
5/16in
# 18

0-0.5 bar

6

6

4

2

2

1

1

0.5-2 bar

6

6

4

2

2

1

1

313D, 314D, STA-PURE, CHEM-SURE to 400 rpm

1.6mm
1/16in
# 14

3.2mm
1/8in
# 16

4.8mm
3/16in
# 25

6.4mm
1/4in
# 17

8.0mm
5/16in
# 18

0-0.5 bar

1

1

1

1

1

0.5-2 bar

1

1

1

1

1

313D2, 314D2, Pumpsil, Marprene, Bioprene, Tygon, Neoprene, Fluorel, STA-PURE, CHEM-SURE to 400 rpm

0.5mm
1/50in
# 112

0.8mm
1/32in
# 13

1.6mm
1/16in
# 14

3.2mm
1/8in
# 16

4.8mm
3/16in
# 25

6.4mm
1/4in
# 17

8.0mm
5/16in
# 18

0-0.5 bar

1

1

1

1

1

1

1

0.5-2 bar

1

1

1

1

1

1

1

23 Tubing for 313D and 314D pumpheads, 1.6mm wall



mm

inch

#

Marprene

Bioprene

Chem-Sure

0.5

1/50"

112

902.0005.016

903.0005.016


0.8

1/32"

13

902.0008.016

903.0008.016


1.6

1/16"

14

902.0016.016

903.0016.016

965.0016.016

3.2

1/8"

16

902.0032.016

903.0032.016

965.0032.016

4.8

3/16"

25

902.0048.016

903.0048.016

965.0048.016

6.4

1/4"

17

902.0064.016

903.0064.016

965.0064.016

8.0

5/16"

18

902.0080.016

903.0080.016

965.0080.016

mm

inch

#

Tygon

Fluorel

Neoprene

0.8

1/32"

13



920.0008.016

1.6

1/16"

14

950.0016.016

970.0016.016

920.0016.016

3.2

1/8"

16

950.0032.016

970.0032.016

920.0032.016

4.8

3/16"

25

950.0048.016

970.0048.016

920.0048.016

6.4

1/4"

17

950.0064.016

970.0064.016

920.0064.016

8.0

5/16"

18

950.0080.016

970.0080.016

920.0080.016

mm

inch

#

Pumpsil

Sta-Pure

Butyl

0.5

1/50"

112

913.A005.016



0.8

1/32"

13

913.A008.016



1.6

1/16"

14

913.A016.016

960.0016.016

930.0016.016

3.2

1/8"

16

913.A032.016

960.0032.016

930.0032.016

4.8

3/16"

25

913.A048.016

960.0048.016

930.0048.016

6.4

1/4"

17

913.A064.016

960.0064.016

930.0064.016

8.0

5/16"

18

913.A080.016

960.0080.016

930.0080.016

Note: 1.6mm wall Chem-Sure and Sta-Pure tubing are supplied in 305mm lengths.

24 Tubing for 313D2 and 314D2 pumpheads, 2.4mm wall



mm

inch

#

Marprene

Bioprene

Pumpsil

0.5

1/50"

 

 

 

913.A005.024

0.8

1/32"

 

 

 

913.A008.024

1.6

1/16"

119

902.0016.024

903.0016.024

913.A016.024

3.2

1/8"

120

902.0032.024

903.0032.024

913.A032.024

4.8

3/16"

15

902.0048.024

903.0048.024

913.A048.024

6.4

1/4"

24

902.0064.024

903.0064.024

913.A064.024

mm

inch

#

Chem-Sure

Sta-Pure


1.6

1/16"

119

965.0016.024

960.0016.024


3.2

1/8"

120

965.0032.024

960.0032.024


4.8

3/16"

15

965.0048.024

960.0048.024


6.4

1/4"

24

965.0064.024

960.0064.024


Note: 2.4mm wall Chem-Sure and Sta-Pure tubing are supplied in 355mm lengths.

25 501RL pumphead

The 501RL and 501RL2 pumpheads are suitable for tubing with internal diameters up to 8.0mm. The 501RL is set during manufacture for use with 1.6mm wall tubing and 501RL2 is set for 2.4mm wall tubing.

The spring-loaded rollers give extended tube life. The pumphead can be run clockwise, for best tube life, or anti-clockwise for higher pressures. The tool-lockable guard should be locked shut while the pump is in use.

25.1 501RL and 501RL2 installation

Always isolate the pump from the mains power supply before opening the guard or performing any positioning, removal or maintenance activity.

The rotor and track may be removed from the pump for cleaning or to reposition the track on the pump in one of three positions.

The rotor grips the drive shaft via a split collet. To remove the rotor, hold it firmly and remove the central rotor retaining screw.

Pull the rotor from the drive shaft, leaving the collet on the drive shaft. If the collet adheres to the rotor, it can be removed by partially re-inserting the rotor retaining screw and tapping it gently. Ensure that the dog within the collet properly re-engages the drive shaft when you reassemble the unit.

The 501RL or 501RL2 track is secured with a locating screw. The track will fit on the drive in three orientations. To change its orientation, remove the track locating screw. Align the track and re-fit the track locating screw. Refit the rotor. Tighten the rotor screw to a torque of 3Nm.

25.2 501RL and 501RL2 tube loading



  • Switch off the mains supply. Unlock and open the pumphead guard.
  • Select a minimum length of 240mm of tubing. Fit one end of the tubing into a clamp.
  • The rotor has tube guide rollers that pull the tube into the pumphead during loading. Turn the rotor carefully until the tube guides pick up the tube. Continue to turn the rotor, and feed the tubing in between the guides.



  • When the tube has gone round the pump track fit the other end of the tubing into the clamp.
  • Check that the tubing has fitted naturally against the track for best tube life. Release the clamps and adjust the tubing if it is slack, twisted, or stretched.
  • The tube clamps can accommodate various tube diameters by pushing in, or pulling out, the grip bars within the clamp. Set the clamps to apply the minimum necessary pressure to the tubing to hold it in position.
  • Restart the pump. Free the downstream clamp for a short time, while the pump is running, so that the tube can find its natural length. Keep fingers clear of the moving rotor. Shut and lock the guard after adjusting the tube.

When using Marprene tubing

Re-tension new tube after the first 30 minutes of running. Stop the pump and release the tube clamp at the outlet of the pump. Pull any slack tube from the pumphead and re-clamp the tube. Re-start the pump. This will correct the normal stretching that occurs with new Marprene tube. The correct tension is essential for good tube life.

25.3 501RL and 501RL2 rotor settings

The 501R and 501R2 pumpheads are factory-set to give optimum tube life with Watson-Marlow tubing. We recommend that the rotors are not adjusted and that other types of tubing are not used.

If the rotor needs re-aligning we recommend the rotor is returned to Watson-Marlow for correct adjustment. Or contact our after-sales department for further information.

Check moving parts of the rotor from time to time for freedom of movement. Lubricate pivot points and rollers occasionally with Teflon lubricating oil.

25.4 501RL and 501RL2 pumphead spares


053.0001.L00

501RL complete pumphead

1

MN0377M

Lockable guard

2

FN4502

Lock

3

FN2341

Hinge screw

4

MN0266M2

Hinge

5

MNA0114A

Tube clamp assembly

6

FN2332

Screw

7

MN0011T

Main roller

8

MNA0143A

501RL rotor assembly

9

SG001

Springs for 501RL (blue)

10

MN0012T

Follower roller

-

XX0095

Teflon lubricant

26 501 Flow rates

Flow rates were obtained using silicone tubing with the pumphead rotating clockwise, pumping water at 20C with zero suction and delivery pressures. For critical applications determine flow rates under operating conditions.

26.1 501RL 1.6mm wall; 501RL2 2.4mm wall: ml/min

501RL, 501RL2

Speed
range

0.5mm
1/50in
# 112

0.8mm
1/32in
# 13

1.6mm
1/16in
# 14

3.2mm
1/8in
# 16

4.8mm
3/16in
# 25

6.4mm
1/4in
# 17

8.0mm
5/16in
# 18

1-300rpm

0.04-13

0.12-37

0.43-130

1.9-560

4.0-1200

6.4-1900

10-3000

27 Tubing for 501RL pumpheads, 1.6mm wall



mm

inch

#

Marprene

Bioprene

Chem-Sure

0.5

1/50"

112

902.0005.016

903.0005.016


0.8

1/32"

13

902.0008.016

903.0008.016


1.6

1/16"

14

902.0016.016

903.0016.016

965.0016.016

3.2

1/8"

16

902.0032.016

903.0032.016

965.0032.016

4.8

3/16"

25

902.0048.016

903.0048.016

965.0048.016

6.4

1/4"

17

902.0064.016

903.0064.016

965.0064.016

8.0

5/16"

18

902.0080.016

903.0080.016

965.0080.016

mm

inch

#

Tygon

Fluorel

Neoprene

0.8

1/32"

13



920.0008.016

1.6

1/16"

14

950.0016.016

970.0016.016

920.0016.016

3.2

1/8"

16

950.0032.016

970.0032.016

920.0032.016

4.8

3/16"

25

950.0048.016

970.0048.016

920.0048.016

6.4

1/4"

17

950.0064.016

970.0064.016

920.0064.016

8.0

5/16"

18

950.0080.016

970.0080.016

920.0080.016

mm

inch

#

Pumpsil

Sta-Pure

Butyl

0.5

1/50"

112

913.A005.016



0.8

1/32"

13

913.A008.016



1.6

1/16"

14

913.A016.016

960.0016.016

930.0016.016

3.2

1/8"

16

913.A032.016

960.0032.016

930.0032.016

4.8

3/16"

25

913.A048.016

960.0048.016

930.0048.016

6.4

1/4"

17

913.A064.016

960.0064.016

930.0064.016

8.0

5/16"

18

913.A080.016

960.0080.016

930.0080.016

Note: 1.6mm wall Chem-Sure and Sta-Pure tubing are supplied in 305mm lengths.

28 Trademarks

Watson-Marlow, Bioprene, Pumpsil and Marprene are trademarks of Watson-Marlow Limited.
Tygon is a trademark of the Saint Gobain Performance Plastics Company.
Fluorel is a trademark of 3M.
Sta-Pure and Chem-Sure are trademarks of W.L.Gore and Associates.

29 Warning not to use pumps in patient-connected applications

Warning These products are not designed for use in, and should not be used for patient-connected applications.

30 Publication history

m-323dz-gb-03.htm: Watson-Marlow 323Dz
First published 12 03.

In compliance with the UK Health and Safety at Work Act and the Control of Substances Hazardous to Health Regulations, you are required to declare the substances which have been in contact with product(s) you return to Watson-Marlow or its subsidiaries or distributors. Failure to do so will cause delays. Please ensure that you fax us this form and receive an RGA (Returned Goods Authorisation) before you despatch the product(s). A copy of this form must be attached to the outside of the packaging containing the product(s). Please complete a separate decontamination certificate for each product.
You are responsible for cleaning and decontaminating the product(s) before return.

Your name

Company

Address

Postcode/zip

Country

Telephone

Fax

Product type

Serial number

To speed the repair, please describe all known faults

The product has ...

Been used Not been used

If the product has been used, please complete all the following sections. If the product has not been used, please just sign this form.

Names of chemicals handled with product(s)

Precautions to be taken in handling these chemicals

Action to be taken in the event of human contact


I understand that the personal data collected will be kept confidentially in accordance with the UK Data Protection Act 1998.

Signature

RGA number

Your position

Date

Please print out, sign and fax to Watson-Marlow Pumps at +44 1326 376009.